Calibration · Free Template

RTD and Thermocouple Sensor Calibration Report Template

Compare a temperature sensor with a reference thermometer in a dry block or bath at several temperatures. The sheet calculates the error in °C and checks it against the tolerance you set.

Excel (.xlsx)Printable PDFFilled exampleFree, no sign up

Download

  • How to Use sheet with steps
  • Template sheet, blank and ready
  • Example sheet, filled in
  • 28 built in formulas
  • A4 landscape print setup
  • Works in Excel, Google Sheets and LibreOffice

Preview of the Filled Example

This is the Example sheet from the download, so you can see exactly how the completed rtd and thermocouple sensor calibration report looks.

Tag numberTE 2205
Sensor typePt100 class A, 3 wire
Use range0 to 300 °C
Tolerance (± °C)0.5
Temperature sourceDry block calibrator
Reference thermometerReference PRT, certificate CC 24610
Calibration date25 Oct 2026
Immersion depth (mm)150

Calibration results

PointSet temperature (°C)Reference (°C)Sensor as found (°C)As found error (°C)As found resultRaw value (Ω or mV)Stable (yes/no)Remarks
10.00.020.100.08PASS100.040Yes
2100.0100.05100.210.16PASS138.590Yes
3200.0199.96200.300.34PASS175.980Yes
4300.0300.08300.520.44PASS212.180YesNear limit
Maximum error (°C):0.44  ·  Overall result:PASS
Advertisement

About the RTD and Thermocouple Sensor Calibration Report

Temperature sensors cannot be adjusted, only checked. A comparison calibration against a reference thermometer in a dry block or bath shows whether the sensor still meets its class, and records raw values for diagnosis.

How to fill in this template

  1. Enter the sensor type and tolerance, for example IEC 60751 class A at the highest test temperature.
  2. Insert sensor and reference in the dry block at the same depth.
  3. At each set point wait for stability, then record reference and sensor readings.
  4. Record the raw Ω or mV value to help diagnose faults.
  5. Review errors and decide: accept, correct in transmitter, or replace.

What the formulas do

  • Error (°C) = sensor reading − reference reading.
  • PASS when |error| ≤ tolerance.

Related tools

Advertisement

Frequently Asked Questions

What tolerance should I use for a class A Pt100?

± (0.15 + 0.002 |t|) °C, so ± 0.35 °C at 100 °C and ± 0.75 °C at 300 °C.

Why use a reference thermometer instead of the dry block display?

The block display measures the block, not the hole the sensor is in; a reference reduces uncertainty.

Can I calibrate a thermocouple the same way?

Yes, set the tolerance from the IEC 60584 class and record mV as the raw value.