What is pH Temperature Compensation?
pH temperature compensation corrects the electrode slope for sample temperature, so the analyser converts mV to pH correctly.
pH Temperature Compensation Explained
Automatic compensation uses a temperature sensor in the probe. It corrects the Nernst slope only, not the real change in sample pH with temperature.
Some analysers add solution compensation for known samples such as pure water.
pH Temperature Compensation Formula
| E | electrode voltage (mV), offset corrected |
| T | temperature (K) |
Calculators for pH Temperature Compensation
Free Downloads
Frequently Asked Questions
What is pH Temperature Compensation?
pH temperature compensation corrects the electrode slope for sample temperature, so the analyser converts mV to pH correctly.
Does compensation give pH at 25 °C?
Not by default; it gives the true pH at the sample temperature.
What if the temperature sensor fails?
The analyser uses a wrong slope; errors grow away from pH 7.
Learn more on the blog
Electrical conductivity explained
A practical guide on instrumentationblog.in with more detail on pH temperature compensation and how it is applied on real plants.
Read Full ArticleRelated Terms
More Analytical Terms
All analytical terms · Full glossary A to Z