What is Sample Lag Time?
transport delayAnalytical
Sample lag time is the delay for a sample to travel from the process tap to the analyser.
Advertisement
Sample Lag Time Explained
Long lines, large tubing and low flow add delay that hurts control and alarms. Fast loops and short lines reduce it.
For gases, compressibility means the delay depends on pressure along the line.
Sample Lag Time Formula
t = V ÷ Q = π d² L ÷ (4 Q)
| d | tube inside diameter |
| L | length |
| Q | flow |
Example: 4 mm ID, 30 m at 1 L/min: V ≈ 0.38 L, t ≈ 23 s.
Advertisement
Calculators for Sample Lag Time
Free Downloads
Frequently Asked Questions
What is Sample Lag Time?
Sample lag time is the delay for a sample to travel from the process tap to the analyser.
What lag is acceptable?
Often under about 60 s for control analysers; set by process needs.
Do filters add lag?
Yes, their internal volume adds delay and mixing.
You May Also Like
Learn more on the blog
Instrumentation and control standards
A practical guide on instrumentationblog.in with more detail on sample lag time and how it is applied on real plants.
Read Full ArticleRelated Terms
Analytical termSample Conditioning SystemA sample conditioning system takes a process sample and makes it suitable for the analyser: filtered,…Temperature termResponse TimeResponse time is how fast a temperature sensor follows a step change, given as a time constant τ (63 %)…Analytical termAnalyser ShelterAn analyser shelter is a ventilated building or cabinet that protects process analysers and their sample…
More Analytical Terms
Analytical termGas ChromatographA gas chromatograph separates a gas or vaporised liquid sample into components in a column and measures…Analytical termNDIR Gas AnalyserAn NDIR gas analyser measures gases such as CO₂, CO and methane from how much infrared light they absorb…Analytical termpHpH is a measure of how acidic or alkaline a water based solution is, based on the hydrogen ion activity.Analytical termpH ElectrodeA pH electrode is a sensor with a pH sensitive glass membrane and a reference electrode, producing a…Analytical termReference ElectrodeA reference electrode provides a stable voltage against which the pH or ORP measuring electrode is compared.Analytical termNernst SlopeNernst slope is the theoretical change in electrode voltage per pH unit, 59.16 mV at 25 °C, rising with…
All analytical terms · Full glossary A to Z
Advertisement