DP Flow Measurement Uncertainty Calculator (ISO 5167)

Find the uncertainty of an orifice, nozzle or venturi flow measurement by combining the uncertainties of each input with the sensitivity coefficients given in ISO 5167-1, and see which input dominates.

ISO 5167-1Uncertainty budgetOrifice and venturiSensitivity
β
± %
± %
± %
± %
± %
± %
Flow uncertainty± 0.602 % of rate
Largest contributorC
How this result was calculated
  1. (δq/q)² = (δC/C)² + (δε/ε)² + [2/(1 − β⁴)]²(δd/d)² + [2β⁴/(1 − β⁴)]²(δD/D)² + ¼(δΔP/ΔP)² + ¼(δρ/ρ)²
  2. δq/q = ± 0.602 %

Uncertainty budget

InputInput uncertaintySensitivityContributionShare of variance
C0.5 %1.0000.5 %69 %
ε0 %1.0000 %0 %
Bore d0.07 %2.2980.161 %7.14 %
Pipe D0.4 %0.2980.119 %3.91 %
ΔP0.5 %0.5000.25 %17.2 %
Density0.2 %0.5000.1 %2.76 %
Advertisement

ISO 5167-1 Uncertainty Equation

(δqm/qm)² = (δC/C)² + (δε/ε)² + [2/(1 − β⁴)]² (δd/d)² + [2β⁴/(1 − β⁴)]² (δD/D)² + ¼(δΔP/ΔP)² + ¼(δρ/ρ)²

Each δx/x is the relative uncertainty of an input at the same confidence level, and the terms are combined in quadrature as ISO 5167-1 describes. The C uncertainty comes from the relevant part of ISO 5167 (about 0.5 % for orifice plates with β from 0.2 to 0.6). The DP uncertainty is that of the transmitter at the operating DP, which grows at low flow. Extra uncertainty for short straight lengths is added arithmetically, not in quadrature.

Worked example

An orifice with β = 0.6, C ± 0.5 %, bore ± 0.07 %, pipe ± 0.4 %, DP ± 0.5 % and density ± 0.2 % gives about ± 0.60 % of flow; C is the largest contributor.

Key insight: DP enters with a factor of ½ because flow goes with √DP, but the DP uncertainty itself rises fast at low flow (a fixed transmitter error is a larger fraction of a small DP). Check uncertainty at minimum flow, not only at maximum.
Advertisement

Frequently Asked Questions

Why does density count only half?

Mass flow is proportional to √(ΔP × ρ), so a relative error in ρ gives half that relative error in flow.

Why is pipe diameter less important than bore?

Its sensitivity is 2β⁴/(1 − β⁴), small at moderate β, while the bore has 2/(1 − β⁴), more than 2.

Is this the total measurement uncertainty?

It covers the primary element and inputs. Add installation effects and, for gases, pressure and temperature uncertainty in density.