Ultrasonic Flowmeter Fault Finder

Choose the ultrasonic meter symptom and get the likely causes for transit time and Doppler meters, with checks and fixes.

No signalWrong readingUnstableZero offset
SymptomNo signal or low signal strength
Most likely causePoor coupling (gel dried or missing)
First checkRemove sensors and inspect

Causes in order of likelihood

CauseHow to checkFix
Poor coupling (gel dried or missing)Remove sensors and inspectReapply couplant; use permanent pads
Wrong sensor spacingCompare with spacing calculated by the meterRemount at correct spacing
Air bubbles or solids scatter the soundCheck processMove to a better location; Doppler for dirty liquids
Advertisement

Before You Start

  • Work under a permit, inform the control room and put the loop in manual or bypass the trip before touching a live instrument.
  • Check signal strength and quality diagnostics.
  • Confirm pipe OD, wall thickness, material and liner in the setup.
  • Confirm the pipe is full at the sensor location.

Ultrasonic Flowmeter Fault Finder: All Symptoms

No signal or low signal strength

Likely causeHow to checkFix
Poor coupling (gel dried or missing)Remove sensors and inspectReapply couplant; use permanent pads
Wrong sensor spacingCompare with spacing calculated by the meterRemount at correct spacing
Air bubbles or solids scatter the soundCheck processMove to a better location; Doppler for dirty liquids

Reading consistently wrong

Likely causeHow to checkFix
Wrong pipe data (OD, wall, liner)Measure wall with a thickness gaugeCorrect pipe data
Short straight runCheck upstream disturbancesMove sensors; use two paths
Pipe partly fullCheck location and fillInstall where pipe is always full

Unstable reading

Likely causeHow to checkFix
Low flow near cutoffCompare with flow rangeAdjust damping or cutoff
Changing gas contentSignal quality variesRemove air; relocate
Temperature change of couplant or pipeCorrelate with temperatureUse suitable couplant and mounting

Shows flow at zero

Likely causeHow to checkFix
Zero offset from mountingStop flow and readRun zero function with full, still pipe
Convection in large hot pipesCheck temperature differencesSet low flow cutoff
Valve passingIsolateRepair valve
Key insight: most clamp on errors are setup errors: a wrong wall thickness changes the sound path and the flow area at the same time.
Advertisement

Frequently Asked Questions

What couplant should I use?

The maker gel for temporary use; permanent pads or epoxy for fixed installations.

Transit time or Doppler?

Transit time for clean liquids, Doppler only with enough particles or bubbles.

How much straight pipe?

Often about 10 D upstream for a single path meter; follow the maker.