Temperature Measurement Cheat Sheet
RTD and thermocouple essentials on one page: conversions, Pt100 resistance values and classes, the Callendar–Van Dusen equation, thermocouple types, millivolts and colour codes.
Download
A one page A4 sheet to print, pin up in the workshop or keep on your phone. Everything on it is also shown below.
Download PDF A4 · 87 KBTemperature units
| Fahrenheit | °F = °C × 9 ÷ 5 + 32 |
| Kelvin | K = °C + 273.15 |
| Rankine | °R = °F + 459.67 |
| Interval | 1 °C (or 1 K) = 1.8 °F |
Pt100 resistance (IEC 60751)
| °C | Ω | °C | Ω |
|---|---|---|---|
| −50 | 80.31 | 200 | 175.86 |
| 0 | 100.00 | 300 | 212.05 |
| 50 | 119.40 | 400 | 247.09 |
| 100 | 138.51 | 500 | 280.98 |
| 150 | 157.33 | 600 | 313.71 |
Pt1000: multiply by 10. α = 0.00385 Ω/Ω/°C, about 0.385 Ω per °C for Pt100.
Callendar–Van Dusen
| 0 °C and above | R = R0 (1 + A t + B t²) |
| Below 0 °C | R = R0 [1 + A t + B t² + C (t − 100) t³] |
| Constants | A = 3.9083 × 10⁻³, B = −5.775 × 10⁻⁷, C = −4.183 × 10⁻¹² |
RTD tolerance classes (IEC 60751)
| Class | Tolerance (°C) |
|---|---|
| AA | ± (0.10 + 0.0017 |t|) |
| A | ± (0.15 + 0.002 |t|) |
| B | ± (0.30 + 0.005 |t|) |
| C | ± (0.60 + 0.01 |t|) |
Thermocouple types
| Type | Materials | Typical range °C | mV at 100 °C |
|---|---|---|---|
| K | NiCr / NiAl | −200 to 1200 | 4.096 |
| J | Fe / CuNi | −40 to 750 | 5.269 |
| T | Cu / CuNi | −200 to 350 | 4.279 |
| E | NiCr / CuNi | −200 to 900 | 6.319 |
| N | NiCrSi / NiSi | −200 to 1200 | 2.774 |
| R | Pt13Rh / Pt | 0 to 1600 | 0.647 |
| S | Pt10Rh / Pt | 0 to 1600 | 0.646 |
| B | Pt30Rh / Pt6Rh | 600 to 1700 | 0.033 |
mV values from the IEC 60584-1 / NIST reference tables, cold junction at 0 °C.
Thermocouple colours (positive / negative)
| Type | IEC 60584-3 | ANSI MC96.1 |
|---|---|---|
| K | Green / white | Yellow / red |
| J | Black / white | White / red |
| T | Brown / white | Blue / red |
| E | Violet / white | Purple / red |
| N | Pink / white | Orange / red |
| R, S | Orange / white | Black / red |
| B | Grey / white | Grey / red |
ANSI: the negative wire is always red.
RTD vs thermocouple
- RTD: more accurate and stable, up to about 600 °C
- Thermocouple: wider range (to 1700 °C), faster, rugged, needs cold junction compensation
- Use 3 wire RTD to cancel lead resistance; 4 wire for laboratory accuracy
- Use extension or compensating cable of the same type for thermocouples
Cold junction
| Measured | Emeas = E(Thot) − E(Tcj) |
| So | E(Thot) = Emeas + E(Tcj), then convert to °C |
Related Calculators and Tools
Frequently Asked Questions
What is the resistance of a Pt100 at 0 °C?
100.00 Ω. At 100 °C it is 138.51 Ω.
Which thermocouple is most common?
Type K, for its wide range (about −200 to 1200 °C) and low cost.
Why is the red wire negative on US thermocouples?
ANSI MC96.1 makes the negative conductor red for every type, so the polarity is always easy to identify.