Temperature Measurement Cheat Sheet

RTD and thermocouple essentials on one page: conversions, Pt100 resistance values and classes, the Callendar–Van Dusen equation, thermocouple types, millivolts and colour codes.

One page A4Print ready PDFFree, no sign up

Download

Temperature Measurement Cheat Sheet PDF preview

A one page A4 sheet to print, pin up in the workshop or keep on your phone. Everything on it is also shown below.

Download PDF A4 · 87 KB

Temperature units

Fahrenheit°F = °C × 9 ÷ 5 + 32
KelvinK = °C + 273.15
Rankine°R = °F + 459.67
Interval1 °C (or 1 K) = 1.8 °F

Pt100 resistance (IEC 60751)

°CΩ°CΩ
−5080.31200175.86
0100.00300212.05
50119.40400247.09
100138.51500280.98
150157.33600313.71

Pt1000: multiply by 10. α = 0.00385 Ω/Ω/°C, about 0.385 Ω per °C for Pt100.

Callendar–Van Dusen

0 °C and aboveR = R0 (1 + A t + B t²)
Below 0 °CR = R0 [1 + A t + B t² + C (t − 100) t³]
ConstantsA = 3.9083 × 10⁻³, B = −5.775 × 10⁻⁷, C = −4.183 × 10⁻¹²
Advertisement

RTD tolerance classes (IEC 60751)

ClassTolerance (°C)
AA± (0.10 + 0.0017 |t|)
A± (0.15 + 0.002 |t|)
B± (0.30 + 0.005 |t|)
C± (0.60 + 0.01 |t|)

Thermocouple types

TypeMaterialsTypical range °CmV at 100 °C
KNiCr / NiAl−200 to 12004.096
JFe / CuNi−40 to 7505.269
TCu / CuNi−200 to 3504.279
ENiCr / CuNi−200 to 9006.319
NNiCrSi / NiSi−200 to 12002.774
RPt13Rh / Pt0 to 16000.647
SPt10Rh / Pt0 to 16000.646
BPt30Rh / Pt6Rh600 to 17000.033

mV values from the IEC 60584-1 / NIST reference tables, cold junction at 0 °C.

Thermocouple colours (positive / negative)

TypeIEC 60584-3ANSI MC96.1
KGreen / whiteYellow / red
JBlack / whiteWhite / red
TBrown / whiteBlue / red
EViolet / whitePurple / red
NPink / whiteOrange / red
R, SOrange / whiteBlack / red
BGrey / whiteGrey / red

ANSI: the negative wire is always red.

RTD vs thermocouple

  • RTD: more accurate and stable, up to about 600 °C
  • Thermocouple: wider range (to 1700 °C), faster, rugged, needs cold junction compensation
  • Use 3 wire RTD to cancel lead resistance; 4 wire for laboratory accuracy
  • Use extension or compensating cable of the same type for thermocouples

Cold junction

MeasuredEmeas = E(Thot) − E(Tcj)
SoE(Thot) = Emeas + E(Tcj), then convert to °C

Related Calculators and Tools

Advertisement

Frequently Asked Questions

What is the resistance of a Pt100 at 0 °C?

100.00 Ω. At 100 °C it is 138.51 Ω.

Which thermocouple is most common?

Type K, for its wide range (about −200 to 1200 °C) and low cost.

Why is the red wire negative on US thermocouples?

ANSI MC96.1 makes the negative conductor red for every type, so the polarity is always easy to identify.