Magnetic Flowmeter Fault Finding Guide

A one page guide to magmeter problems: grounding, empty pipe, electrode coating, conductivity limits and configuration.

One page A4Vendor neutralFree, no sign up

Download

Magnetic Flowmeter Fault Finding Guide PDF preview

A one page A4 guide to print for the workshop or keep on your phone in the field. The full content is also shown below.

Download PDF A4 · 55 KB

Before you start

  • Permit, isolation and control room informed.
  • Put the loop in manual or inhibit trips as authorised.
  • Use certified test equipment in hazardous areas.
  • Record as found readings before changing anything.

Quick checks

  • Pipe full at the meter?
  • Grounding rings or electrodes connected?
  • Conductivity above the meter minimum?
  • Sensor factor matches tag?
Advertisement

Symptoms, causes and fixes

SymptomLikely causesChecks and fixes
Reading zero with flowEmpty pipe detection active, coil fault, wrong configurationCheck pipe full; coil resistance; flow direction setting
Noisy or jumpingPoor grounding, air bubbles, solids hitting electrodes, low conductivityFit grounding rings; remove air; check conductivity
Reads high or lowWrong size or calibration factor, partly full pipe, coatingCheck sensor factor and DN; check fill; clean electrodes
Reads flow when stoppedElectrode coating, grounding, valve passingClean electrodes; check grounding; confirm valves
Empty pipe alarm with full pipeCoated electrodes, very low conductivity, wrong thresholdClean; check conductivity; adjust threshold
Negative flowReverse direction setting or installed backwardsCheck arrow and direction setting
Advertisement

Installation reminders

  • Vertical up flow or horizontal with electrodes horizontal.
  • Not at the top of a pipe or before a free outlet.
  • About 5 D upstream, 2 D downstream (typical).
  • Separate coil and signal cables from power cables.

NAMUR NE107 status

StatusMeaningAction
F FailureOutput invalidRepair or replace; check loop goes to safe state
C Function checkSimulation, calibration or maintenance activeFinish the work, return to normal mode
S Out of specificationOperating outside limits or uncertainCheck process conditions and sensor
M Maintenance requiredStill valid but degradingPlan maintenance before failure

Related Tools, Procedures and Calculators

Frequently Asked Questions

Why must magmeters be grounded?

The liquid potential must be tied to the meter reference; otherwise noise swamps the tiny flow signal.

What conductivity is needed?

Commonly at least about 5 µS/cm for standard meters; check the maker data.

Can a magmeter be verified in place?

Many have built in verification of coils and electronics.