Fault Finding Guides · Free PDF
Coriolis Flowmeter Fault Finding Guide
A one page guide to coriolis meter problems: zero stability, gas or solids in liquids, drive gain, density and mounting stress.
One page A4Vendor neutralFree, no sign up
Download
A one page A4 guide to print for the workshop or keep on your phone in the field. The full content is also shown below.
Download PDF A4 · 53 KBBefore you start
- Permit, isolation and control room informed.
- Put the loop in manual or inhibit trips as authorised.
- Use certified test equipment in hazardous areas.
- Record as found readings before changing anything.
Quick checks
- Meter full of single phase fluid?
- Drive gain normal compared with commissioning?
- Zero recorded at process conditions?
- No pipe stress on the meter?
Advertisement
Symptoms, causes and fixes
| Symptom | Likely causes | Checks and fixes |
|---|---|---|
| Flow at zero | Zero not set at process conditions, mounting stress, valve passing | Zero with line full and still; check supports; confirm valves |
| Noisy flow and density | Gas bubbles or solids (two phase), cavitation | Raise pressure; check drive gain; locate meter lower in line |
| Density wrong | Coating, erosion, gas entrainment, wrong calibration | Check with lab sample; clean or check tubes |
| High drive gain | Two phase flow, coating, tube damage | Remove gas; inspect; contact maker if persistent |
| Reading shifts with temperature | Zero not done at operating temperature, stress | Re zero hot; check pipe expansion and supports |
| Sensor fault alarm | Pickoff or drive coil failure, cable fault | Check cable and terminal resistance per manual |
Advertisement
NAMUR NE107 status
| Status | Meaning | Action |
|---|---|---|
| F Failure | Output invalid | Repair or replace; check loop goes to safe state |
| C Function check | Simulation, calibration or maintenance active | Finish the work, return to normal mode |
| S Out of specification | Operating outside limits or uncertain | Check process conditions and sensor |
| M Maintenance required | Still valid but degrading | Plan maintenance before failure |
Related Tools, Procedures and Calculators
Frequently Asked Questions
When should a coriolis meter be zeroed?
At commissioning and after changes, with the meter full, flow stopped and at process temperature and pressure.
What does high drive gain mean?
The meter works harder to vibrate the tubes, often because of gas or coating.
Does coriolis need straight pipe?
Usually not, but supports and no stress are important.
You May Also Like
Learn more on the blog
Coriolis flow meter uncertainty
A practical guide on instrumentationblog.in with more background on this equipment.
Read Full ArticleMore Fault Finding Guides
Pressure Transmitter Fault Finding GuidePrintable A4 PDFDP Flow Measurement Fault Finding GuidePrintable A4 PDFMagnetic Flowmeter Fault Finding GuidePrintable A4 PDFVortex and Ultrasonic Flowmeter Fault Finding GuidePrintable A4 PDFRadar and Ultrasonic Level Fault Finding GuidePrintable A4 PDFDP Level and Guided Wave Radar Fault Finding GuidePrintable A4 PDFTemperature Measurement Fault Finding GuidePrintable A4 PDFControl Valve and Positioner Fault Finding GuidePrintable A4 PDFOn Off and ESD Valve Fault Finding GuidePrintable A4 PDFpH and Conductivity Analyser Fault Finding GuidePrintable A4 PDFFixed Gas Detector Fault Finding GuidePrintable A4 PDFPLC Analog Input Fault Finding GuidePrintable A4 PDFModbus and HART Communication Fault Finding GuidePrintable A4 PDFVFD Fault Finding Guide (Vendor Neutral)Printable A4 PDFNAMUR NE107 Diagnostic Status GuidePrintable A4 PDF