Free engineering calculators and unit convertersceo@instrumentationblog.in
Calibration · Free Template

Flowmeter Verification Record Template

Record a flowmeter check against a reference (master meter, prover, weigh tank or bucket test) at several flow rates. The sheet calculates the error at each run and the mean error per flow rate.

Excel (.xlsx)Printable PDFFilled exampleFree, no sign up

Download

  • How to Use sheet with steps
  • Template sheet, blank and ready
  • Example sheet, filled in
  • 54 built in formulas
  • A4 landscape print setup
  • Works in Excel, Google Sheets and LibreOffice

Preview of the Filled Example

This is the Example sheet from the download, so you can see exactly how the completed flowmeter verification record looks.

Tag numberFT 4102
Meter typeMagnetic flow meter DN 80
ReferenceWeigh tank with calibrated scale
Volume unitL
Tolerance (± % of reading)0.5
Test date25 Oct 2026

Verification runs

RunNominal flow (% of max)Flow rate (m³/h)Duration (s)Reference volumeMeter volumeError (% of reading)ResultByRemarks
125 %12.50120416.20417.300.26PASSTech A
225 %12.50120415.80417.000.29PASSTech A
350 %25.0090624.50625.900.22PASSTech A
450 %25.0090625.10626.200.18PASSTech A
5100 %50.0060832.00834.100.25PASSTech A
6100 %50.0060833.40835.200.22PASSTech A
Mean error (%):0.24  ·  Spread of errors (max − min, %):0.11  ·  Runs failed:0
Advertisement

About the Flowmeter Verification Record

Flowmeters drift through wear, coating and electronics ageing. Periodic verification against a trusted reference shows whether the meter still meets its accuracy, and provides data for a meter factor correction if needed.

How to fill in this template

  1. Enter meter, reference and tolerance.
  2. Run the flow steady at each test rate and collect a reference volume while reading the meter total.
  3. Enter reference and meter volumes for each run.
  4. Review error, mean error and spread; a consistent offset suggests a K factor or meter factor correction.
  5. Sign and file with the meter history.

What the formulas do

  • Error (%) = (meter volume − reference volume) ÷ reference volume × 100.
  • Mean error = average of all run errors.
  • Spread = maximum − minimum error, a simple repeatability check.

Related tools

Advertisement

Frequently Asked Questions

How long should each run be?

Long enough that timing and reading resolution are small compared with the tolerance, often at least 1 minute or 1000 pulses.

What is a good reference?

A weigh tank, a prover or a master meter with uncertainty at least 3 to 4 times better than the tolerance.

Can I use a bucket test?

For small water meters, yes, with a calibrated container and stopwatch.

You May Also Like

Learn more on the blog

What factors affect the stability of flow meter readings?

A practical guide on instrumentationblog.in that explains the procedure behind this template and how engineers apply it on real plants.

Read Full Article

More Calibration Templates

All Templates and Downloads