Temperature Transmitter Calibration Report Template

A calibration report for RTD and thermocouple temperature transmitters. Record the simulated sensor signal at each point; the sheet works out the expected mA, error in % of span and result.

Excel (.xlsx)Printable PDFFilled exampleFree, no sign up

Download

  • How to Use sheet with steps
  • Template sheet, blank and ready
  • Example sheet, filled in
  • 66 built in formulas
  • A4 landscape print setup
  • Works in Excel, Google Sheets and LibreOffice

Preview of the Filled Example

This is the Example sheet from the download, so you can see exactly how the completed temperature transmitter calibration report looks.

Tag numberTT 2101
ServiceReactor R 201 jacket outlet
Make and modelYokogawa YTA110
Serial numberS5N812443
Sensor typePt100, 3 wire (IEC 60751)
Engineering unit°C
Calibrated range LRV0
Calibrated range URV200
Tolerance (± % of span)0.25
Damping (s)2
Simulator / calibratorFluke 726
Standard certificate noNABL/CC 24203
Calibration date12 Oct 2026
Ambient temperature (°C)27

Calibration results (sensor simulated at the transmitter input)

Test point (% of span)Applied temperatureSimulated signal (Ω or mV)Expected output (mA)As found output (mA)As found error (% span)As found resultAs left output (mA)As left error (% span)As left result
00.0100.004.0004.0060.04PASS4.0010.01PASS
2550.0119.408.0008.0120.08PASS8.0020.01PASS
50100.0138.5112.00012.0200.12PASS12.0010.01PASS
75150.0157.3316.00016.0260.16PASS16.0000.00PASS
100200.0175.8620.00020.0300.19PASS20.0020.01PASS
Maximum as found error (% span):0.19  ·  Maximum as left error (% span):0.01  ·  Overall result (as left if adjusted, else as found):PASS
Advertisement

About the Temperature Transmitter Calibration Report

Temperature transmitters convert an RTD resistance or thermocouple millivolt signal to 4 to 20 mA. The transmitter is normally calibrated by simulating the sensor with a calibrator, so the report records both the simulated signal and the output. The sensor itself is checked separately in a dry block or bath.

How to fill in this template

  1. Enter the tag, sensor type and calibrated range (LRV and URV in °C) and the tolerance.
  2. Disconnect the sensor and connect the calibrator in RTD (Ω) or thermocouple (mV) simulation mode.
  3. For each test point simulate the Applied temperature, write the simulated Ω or mV in the yellow column and the transmitter output in As found output (mA).
  4. Adjust (sensor trim or output trim) if any point fails and record the As left readings.
  5. Reconnect the sensor and confirm a sensible process reading before closing the work order.

What the formulas do

  • Applied temperature = LRV + (URV − LRV) × test point % ÷ 100.
  • Expected output = 4 + 16 × test point % ÷ 100 mA.
  • Error (% of span) = (measured mA − expected mA) ÷ 16 × 100; PASS when within the tolerance.

Related tools

Advertisement

Frequently Asked Questions

Should I calibrate the sensor and transmitter together?

A loop calibration with the real sensor in a dry block or bath checks both together and gives the best total accuracy. Simulating the sensor checks only the transmitter.

What resistance does a Pt100 have at 100 °C?

138.51 Ω according to IEC 60751.

Why record the simulated value?

It proves which input was applied, so anyone can repeat or audit the test.